Fx Calibration
Manual push using the 250N calibration sensor recently calibrated
Calibration on the Fx support
Unit Range (Right push @47N/ 0 after preload): 505N (far from load cell rating 75lbs = 333.7N)
Applied load (+-1N) | Right Compression (0 after preload >50N) | Left Compression (0 after preload >50N) |
20N | 20N | 22N |
30N | 30N | 32N |
50N | 50N | 50N |
Calibration on the ball
Unit Range (Right push @47N/ 0 after preload): 370N (closer to load cell rating 75lbs = 333.7N)
Applied load (+-1N) | Right Compression (0 after preload >50N) | Left Compression (0 after preload >50N) |
20N | 19N | 19N |
30N | 29N | 31N |
50N | 51N | 54N |
Return to 0 after compression
Right Compression, Bias after preload: (+-1N) | Left Compression reading after load: (+-1N) |
5N | -2.2N |
10N | -2.2N |
30N | -6.7N |
50N | -6.5N |